Original Article
Effect
of Contact Lens Wear on Tear Film Break up Time (TBUT) among Contact Lens Users
Rabia Ammer
Pak J Ophthalmol 2017, Vol. 33, No. 3
. . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .
.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
See end of article for authors affiliations …..……………………….. Correspondence to: Rabia
Ammer, Department
of Allied Health Sciences School of
Optometry/ The University of Faisalabad Email: rabbia.ammer@gmail.com |
Purpose: To
determine the effect of contact lens wear on tear film break up time (TBUT) and to find out correlation between tear
film break up time (TBUT) with type of contact lens wear, daily wearing time
of contact lens, years of contact lens use and power of contact lens. Study
Design: Cross-sectional study and convenient sampling was used. Place
and Duration of Study: This study recruited the
sample from Madinah Teaching Hospital Faisalabad, E Plomer Optics and Punjab
Optics Lahore. The study was conducted in 4 months from 05-04-2016 to
05-08-2016. Materials
and Methods: TBUT determined by using slit lamp and
fluorescein strips. SPSS version 23 was used for data analysis. Inferential
statistics was reported for variables. Results:
A sample of 100 (67 females and 33 males) contact lens users
recruited for the study. Mean age of sample was 30.10 ± 7.86 years. Mean
daily wearing time of contact lens was 9.82 ± 2.19 hours/day and mean of
years of contact lens use was 8.35 ± 5.81 years. 67% of 100 contact lens
users had abnormal (less than 10 sec.) tear film break up
time (TBUT). A significant negative
correlation of TBUT was found with daily wearing time of contact lenses (r = -.251), years of contact lens use (r = -.542) and minus power of contact lens (r=-.330). No significant association of TBUT was
found with type of contact lens and plus power of contact lens. Conclusions: TBUT decreases with increase in daily wearing time,
years of contact lens use and high minus power of contact lens, while type of
contact lens and plus power of contact did not significantly affect the TBUT. Key
Words: contact lens, tear film break up time, TBUT, tear film stability. |
The ocular surface consists of various glandular tissues, which
secrete the tear film that coats and protect the ocular surface1. The
tears are distributed by normal, non-voluntary action of eyelid and each blink
refreshes the pre-corneal tear film. When blinking is held, evaporation of
aqueous layer of tear film started which causes thinning of the tear film in a
localized area. Subsequently dry spots are formed in tear film when the tears
evaporate2. Tear film stability is usually measured by its lack of
stability, by a test called Tear film break-up time (TBUT)1. TBUT is
the time taken in seconds for the tear film to break following a blink
cessation2. Tear break up time TBUT of 15-45 sec is considered
normal while less than 10 sec is abnormal2 and less than 5 sec is
suggestive of symptoms of dry eye disease 1.
The
physical placement of a contact lens over the ocular surface divides the tear
film into two parts, i.e. a pre-lens tear film and a post-lens tear film and it
forms a new interface within the ocular atmosphere. This division and new
interface induce changes in biophysical properties of tear film3.
When a contact lens is placed on the eye, it affects the stability of tear
film, evaporation rate of tears, thickness of lipid layer and volume of tears.
The contact lens use mainly alters the normal sequence of tear film function
and cause distractions in quantity and quality of the tear film, which lead to
contact lens intolerance4.Both RGP and soft contact lenses interfere
with the tear film stability and decreases the TBUT. RGP contact lenses reduce
TBUT up to 4 to 6 sec whereas soft contact lenses decrease TBUT up to 4 to 10
sec5.This study is conducted to determine the effect of
contact lens wear on tear film break up time (TBUT)
and to find out correlation between tear film break up time (TBUT) with type of
contact lens wear, daily wearing time of contact lens, years of contact lens
use and power of contact lens.
MATERIAL AND METHODS:
It was
a cross-sectional study and convenient sampling technique was used to collect
the sample of 100 contact lens users. The study was conducted in 4 months from
05-04-2016 to 05-08-2016. Data were collected from 3 different settings;
Madinah Teaching Hospital Faisalabad, E Plomer Optics and Punjab Optics,
Lahore. For
ethical concerns, approval of study obtained from the ethical review board of
The University of Faisalabad in accordance with the principles of Declaration
of Helsinki. Subjects aged 15 to 55 years, those used
contact lenses for more than 1 year and without any complain / symptom related
to contact lens use were included in the study. Subjects suffering from any
disease of the cornea or conjunctiva and those used contact lenses for less
than 1 year, excluded from the study. An informed consent form delivered
to gain consent from participants for their voluntary participation by briefly
describing the study topic, its purpose, duration and assuring for
confidentiality of respondents personal information. Subject’s
demographic details, history related to the type, daily wearing time, power and
years of contact lens use were recorded in specially designed self-structured
Performa. A detailed slit lamp examination was performed in a consistent,
orderly fashion from eyelid to cornea to determine any ocular pathology. To measure
tear film break-up time (TBUT), the subject was asked to look upward and sodium
fluorescein was applied to the sclera at lower fornix by using fluorescein
strip moistened by normal saline and then subject was asked to blink several
times. After that subject was instructed to avoid blinking and the tear film
was observed between blinks with a slit lamp under cobalt blue light and time
noted between a complete blink and the appearance of the first black dry spot.
SPSS version 23 was used for data analysis. Descriptive and Inferential
statistics were generated and reported for variables.
RESULTS
A sample of 100 contact lens users recruited in which female
contact lens users were 67 (67%) and male contact lens users were 33 (33%). Age
of contact lens users ranged from 16 to 55 years with mean age of 30.10 ± 7.86
years. The study subjects were wearing different types of contact lenses. 54%
of subjects were soft contact lens users, 17% were soft cosmetic contact lens
users, 14% were RGP contact lens users, 12% were soft toric contact lens users
and 3% were silicone hydrogel contact lens users. The daily wearing time of
contact lens determined in this study ranged from 4 to 16 hours/day and mean
value was 9.82 ± 2.19 hours/day. The years of contact lens use found in this
study ranged from 1.5 to 30 years with a mean value of 8.35 ± 5.81 years. In
this study very high proportion of contact lens users was myopic (96 %) and
used contact lenses of minus power. Minus power of contact lens ranged from
-0.50 to -17.00 D with a mean value of -4.46 D ±3.69 D. The proportion of
hyperopic contact lens users was less (4 %) and plus power of contact lenses
ranged from +2.00 to +5.00 D with a mean value of +4.00D ±1.35 D.
TBUT value was
ranging from 4 to 18 sec with a mean value of 9.14 ± 2.89
seconds in contact lens users. In males average TBUT was 8.88 ± 2.50 sec and in female it was 9.27 ± 3.08 sec. The results
determined that 67% of contact lens users had abnormal TBUT and 33% had normal
TBUT.
No significant (p =
> 0.05) association was found between TBUT and types of contact lenses
(Table 1).
A significant (p =
.012) negative correlation was found between TBUT and daily wearing time of
contact lenses (Table 2). Regression model shows that with 1 hour increase in
daily wearing time, TBUT decreased by -.332 times (Figure 1).
A highly
significant (p = .000) negative correlation was found between TBUT and years of
contact lens use (Table 2). Regression model shows that with 1 year increase in
contact lens use, TBUT decreased by - .270 times (Figure 2).
Table 1:
Association between TBUT and
types of contact lenses.
Type
of Contact Lens |
TBUT (Sec) |
Total |
|
Abnormal (≤ 10) |
Normal (>10) |
||
Soft |
32 |
22 |
54 |
Hard |
10 |
4 |
14 |
Soft Toric |
8 |
4 |
12 |
Silicone Hydrogel |
3 |
0 |
3 |
Soft Cosmetic |
14 |
3 |
17 |
Total |
67 |
33 |
100 |
Pearson Chi-Square 4.878 p-value .300 |
Table 2: Significant
Correlation of TBUT.
|
Daily Wearing Time of CL (Hours/Day) |
Years of CL Use (Years) |
Minus CL Power (D) |
|
TBUT (sec) |
Pearson Correlation |
-.251* |
-.542** |
-.330** |
Sig. (2-tailed) |
.012 |
.000 |
.001 |
|
N |
100 |
100 |
96 |
|
*. Correlation is
significant at the 0.05 level (2-tailed). **. Correlation is
significant at the 0.01 level (2-tailed). |
A
strong significant (p = .001) negative correlation was found between TBUT and
minus power of contact lens (Table 2). Regression model shows that with -1 D increase in power
of contact lens, TBUT decreased by -.263
times (Figure 3).
No significant (p
> 0.05) association was found between corneal changes and plus power of
contact lens.
DISSCUSSION
In this study a
decreased value of TBUT was found in majority of contact lens users. The results determined that 67% of contact lens users had
abnormal TBUT. The reason might be that the
contact lens induced hypoxia interfered with tear film stability and resulted
in shortened TBUT. These results were similar to the studies of Sweeney et al.1, Craig et al.3, Eghosasere et al.4, Du Toitet al.6, Thai et al.7, Glasson et al.8, Nichols and Sinnot9,
Riley et al.10, Stapleton et al.[11], Janine
et al.12, Guillon and
Maissa13, Jansen et al.14,
Shrestha et al.15, Young et al.16, Kastelanet al.17, Gupta et al.18
and Pili et al.19. However
these findings were contrary to the results reported by Santodomingo-Rubido et al.20 study. This could be
due to the regional, racial or environmental differences.
In this study no
significant association was found between TBUT and type of contact lens. The study results were similar to Craig et al.3 and Thai et al.7 studies. But contrary
to Sweeney et al.1,
Eghosasere
et al.4, Riley et al.10, Kastelan et al.17 studies which found association between TBUT and type of
contact lenses. The reason might be that more than half of total subjects used
soft contact lenses and proportions of other types of contact lens users were
less, therefore, could not find any significant association.
Fig. 1: Regression model of TBUT and daily wearing time of contact
lenses (showed that as the daily wearing time of contact lens increased, the
TBUT value decreased).
Fig. 2: Regression model of TBUT and Years of Contact Lens Use (showed
that as the number of years of contact lens use increased, the TBUT value
decreased).
Fig. 3: Regression model of TBUT and
minus power of contact lenses (showed that as the minus power of contact lens
increased, the TBUT value decreased)
In
this study a significant negative correlation was found between TBUT and daily
wearing time of contact lenses. The reason could be that increased daily
wearing time of contact lens caused more hypoxia which adversely affected the
tear film stability. These results were in line with those reported by Kastelan
et al.17 study.
In
this study a highly significant negative correlation was found between TBUT and
years of contact lens use. This might be due to the reason that prolonged use
of contact lenses resulted in chronic hypoxia which unfavorably affected the
tear film stability. No other study could be found to sufficiently discuss the
results.
In
this study a strong significant negative correlation was found between TBUT and
minus power of contact lens. The reason could be that high power contact lenses
were thicker which reduced oxygen permeability and caused hypoxia which lead to
reduce TBUT. No other study was found to sufficiently discuss the results.
In this study no
significant association was found between corneal changes and plus power of
contact lens. These insignificant results could be due to very less number of
hyperopic contact lens users in the study. No other study was found to sufficiently
discuss the results.
CONCLUSION
It is concluded
that long term use of contact lenses decreased the stability of tear film and
thus TBUT. TBUT was more affected with increase in daily wearing time, years of
contact lens use and high minus power of contact lens.
Author’s Affiliation
Dr. Rabia Ammer
BSc Optom, Mphil
Optom.
Department of
Allied Health Sciences/ School of Optometry/ The University of Faisalabad.
Role of Authors
Dr. Rabia Ammer
Concept, Design of
study, Sample selection. Data collection, Data entry, Data analysis, Critical
review, Drafting and revision of manuscript.
REFERENCES
1.
Sweeney
D, Millar T, Raju S. Tear film
stability: A review. Experimental Eye Research, 2013; 117: 28-38.
2.
Williams L.
Anatomy and physiology of the anterior segment. Module1. Anterior segment of
the eye. The IACLE contact lens course. The International Association of
Contact Lens Educators Sydney, Australia, 2000; 1 (1): 3-80.
3.
Craig PJ, Willcox DPM, Arg¨ueso P, Maissa C, Stahl U, Tomlinson A,
Wang J, Yokoi N, Stapleton F. The
TFOS International Workshop on Contact Lens Discomfort: Report of the Contact
Lens Interactions with the Tear Film Subcommittee. Invest Ophthalmol Vis Sci. 2013; 54 (11): TFOS123.
4.
Eghosasere I, Joy EI, Joy OI. Effect of soft contact lens materials on tear film stability and
central corneal radius of curvature: A comparative study of Polymacon and
Lotrafilcon B. Sierra
Leone J Biomed Res. 2011; 3 (3): 144-150.
5.
Terry R.
Corneal oxygen requirements and the effects of hypoxia. Module 6. The cornea in
contact lens wear. The IACLE contact lens course. The International Association
of Contact Lens Educators Sydney, Australia, 2000; 1 (1): 3-36.
6.
Du Toit R, Situ P, Simpson T, Fonn D. The effects of six months of contact lens wear on the tear film,
ocular surfaces, and symptoms of presbyopes. Optom Vis Sci. 2001; 78 (6): 455-462.
7.
Thai
L, Tomlinson A, DoaneM. Effect of
Contact Lens Materials on Tear Physiology. Optom Vis
Sci. 2004; 81 (3): 194-204.
8.
Glasson M, Stapleton F, Willcox, M. Changes to tear film parameters during wear of HEMA-based hydrogel
lenses. Ocul Surf. 2005; 3 (1): 66.
9.
Nichols JJ, Sinnott TL. Tear film, contact lens, and patient-related factors associated
with contact lens–related dry eye. Invest Ophthalmol Vis Sci. 2006; 47 (4): 1319-1328.
10.
Riley C, Young G, Chalmers R. Prevalence of Ocular Surface Symptoms, Signs, and
Uncomfortable Hours of Wear in Contact Lens Wearers: The Effect of Refitting
with Daily-Wear Silicone Hydrogel Lenses (Senofilcon A). Eye Contact Lens, 2006; 32 (6): 281-286.
11.
Stapleton F, Stretton S, Papas E, Skotnitsky C, Sweeney DF. Silicone hydrogel contact lenses and the ocular surface. Ocul
Surf. 2006; 4 (1): 24-43.
12.
Janine A, Smith AJ, Albeitz J, Begley C, Caffery B, Nichols K,
Schaumberg D, Schein O. The
epidemiology of dry eye disease: report of the epidemiology subcommittee of the
international dry eye work shop. Ocul Surf. 2007; 5 (2): 93-107.
13.
Guillon M, Maissa C.
Contact lens wear affects tear film evaporation. Eye Contact Lens, 2008; 34 (6):
326-330.
14.
Jansen M, Begley C, Himebaugh N, Port N. Effect of contact lens wear and a near task on tear film break-up.
Optom Vis Sci. 2010; 87 (5): 350-357.
15.
Shrestha G, Sujakhu D, Shrestha JB, Shrestha JK. Tear film evaluation in contact lens wearers and non wearers. Journal of Institute of Medicine,
2012; 34 (2): 14-20.
16.
Young G, Chalmers R, Napier L, Kern J, Hunt C, Dumbleton K. Soft contact lens-related dryness with and without clinical
signs. Optom Vis Sci. 2012; 89 (8): 1125-1132.
17.
Kastelan
S, Lukenda A, Salopek-Rabatic J, Pavan J, Gotovac M.
Dry eye symptoms and signs in long-term contact lens wearers. Coll Antropol.
2013; (37) 1: 199–203.
18.
Gupta A, Shah M, Samanta A. Effect of extrinsic controls on blinking and tear film stability
among soft contact lens wearers. International
Journal of Medicine & Health Research, 2014; 1 (1): 1-7.
19.
Pili K, Kaštelan S, Karabatic M, Kasun B, Culig B. Dry eye in contact lens wearers as a growing public health
problem. Psychiat Danub. 2014; 26 (3): 528-532.
20.
Santodomingo-Rubido J, Wolffsohn JS, Gilmartin B. Changes in ocular physiology, tear film characteristics, and
symptomatology with 18 months silicone hydrogel contact lens wear. Optom Vis
Sci. 2006; 83 (2): 73-81.